Technical Specifications:
- SIGNAL PROCESSING ECG analysis frequency: 500 Hz
- ST measurements: ST amplitudes, slope, integral, index, ST/HR slope, ST/HR loops, ST/HR index up to 15 leads
- E, J and post-J point: Manual or computer selected
- Local storage: Minimum 3.2 Gb hard drive storage of complete ECG record and test results
- Display: 19" flat panel color display
- ECG output: Real-time ECG/QRS beep/TTL synchronization output
- Full disclosure ECG: Beat-to-beat ECG record and event review
- ECG interpretation: (Optional) 12SL7trade; adult and pediatric ECG analysis program
- Sampling rate: Over-sampling @ 4000 Hz, 12 leads
- Monitored leads: 3, 6 or 12
- Displayed leads: Number on screen 3 or 6 (12 lead check)
- Display format: 3 rhythm, 3 rhythm + medians, 3 rhythm + trends, 6 rhythm, 4 X 2.5 + 1 rhythm, 2 x 6 Display speeds: 25, 50 mm/s
- Displayed data: ST scan/median complexes, arrhythmias, ventricular ectopic/min counter, 3 to 15 waveforms, lead check torso and 12 leads, waterfall display, trends, tabular summary, stored ECG strips, interpretation, time-of-day clock, patient name, warning messages and prompts
- Writer technology: "Instant" load, thermal dot array
- Writer speeds: 5, 12.5, 25 and 50 mm/sec (+/- 2%)
- Frequency response: -3 dB, display and writer
- High pass filter: 0.01 (or 0.05 Hz, special use) with DC offset control
- Low pass filter: 20, 40, 100, 150 Hz (selectable)
- Line filter: 50.0 or 60.0 Hz notch filter (selectable)
- Power supply: AC operation only
- Operating voltage range: 90-135 VAC, 45-63 Hz, 2.8 A /180-270 VAC, 45-63 Hz, 1.4 A
- Temperature: +10 to +40°C (+50 to 104°F)
- Writer resolution: Horizontal 1000 lines/sec x 200 dpi dedicated local printing. 200 x 200 dpi generic printing
- Paper type: Thermal, perforated, fan fold, 300 sheets/pack
- Height: 160 cm(63 in) with 17" monitor installed
- Width: 76 cm(30 in) with paper tray removed, 104 cm (41 in) with paper tray
- Depth: 86.5 cm (34 in)
- Weight: 68 kg (150 lb) without monitor, 86 kg (190 ob) with 17" monitor
|