Technical Specifications:
- ECG Leads
10 lead 12 channel simulataneous 10-second resting ECG
- Dimension
296 (w) x 305.5 (h) x 92.5 (d) mm Approximately 2.98 kg
- Recording Channel
3,6,12 channel and 1 channel (60 sec)
- Sensitivity
5, 10, 20 auto (I~aVF: 10, V1~V6:5) mm/mV
- Printing Speed
12.5, 25, 50 mm/s
- Filters
AC (50/60 Hz, -20dB or better) Muscle (25~35Hz, -3dB or better) Base Line Drift (0.1Hz, -3dB or better) Low pass filter (off, 40Hz, 100Hz, 150Hz)
- Display
2 x 16 character LCD display
- LCD display LED indicator
Sensitivity, speed, filter status, HR, printing form, rhythm lead signal quality, power source, battery
- Keyboard
Membrane keyboard Alphanumeric and symbol available
- Patient Data
ID, name, age, sex, height, weight
- Basic Measurement
Heart rate, PR, QRS, QT/QTc, P-R-T axis
- Recorder
Thermal print head, thermal roll paper Report papers: -Width: A4: 210mm or 8.5" -Length: A4: 300mm or 11" Effective writing width: 204mm Resolution: - Vertical: 8 dots/mm - Horizontal: 16 dots/mm
- Electrical
Internal noise: 20uV(p-p) max Input circuit: Floating input Isolated and defibrillation protected Input impedance: >10MQ Input range: > +/- 5mV Common Mode Rejection: > +100dB DC offset voltage: > +/- 300mV Time constant: 3.2 sec Patient leakage current: <10uA Frequency response: 0.05~150Hz
- Signal qualilty control
Disconnected lead detection
- Power
Power Supply: AC or built-in Battery (option) Power requirement: 100 - 240 Vac, 50/60Hz, 1.0 - 0.5A Power consumption: 60w max
- Battery
capacity 1 hour of normal use (approximately 100 automatic ECG printouts)
- Communication PC connection with RS-232 interface and LAN
- Safety
Class 1, Type BF Conformity CE, CSA
- Environment
Operating humidity: 30~85% Operating temperature: 10° ~40° C Atmospheric pressure: 70 ~ 106 KPa
- Standard accessory
Power Cable - 1 ea. Patient Cable - 1 ea. Electrode - 10 ea. Chart Paper - 1 ea. Operation Manual - 1 ea.
|