DIMENSIONS:
|
60" x 25" x 34" (152 cm x 63 cm x 86 cm)
|
| |
|
WEIGHT:
|
170 lbs. (78 kg)
|
| |
|
Type:
|
15" High-resolution color with integrated touchscreen;
|
|
Mount:
|
User-adjustable tilt (•15°) and swivel (•90°);
|
|
Content:
|
3, 4 or 6 ECG channels, real-time; information and alphanumeric data;
|
|
Displayed leads:
|
User-selectable during a test; (3-, 4- and 6-channel); Waveform display •8 seconds per lead of ECG in 3-channel mode
|
| |
|
Main keyboard:
|
QWERTY layout;
|
|
Ergometer control:
|
Dedicated keys for ergometer speed, grade, start and stop
|
| |
|
Digital I/O:
|
RS-232 interfaces to treadmill, bicycle ergometer, and optional automatic blood pressure monitor; TTL pulse for QRS detection, 50 msec. delay;
|
|
Analog ouput:
|
4 user-selectable outputs
|
| |
|
Network:
|
Ethernet 10baseT for full disclosure report storage and export of final report data using HL7 protocol or XML;
|
|
Removable media:
|
Zip drive for full-disclosure test storage – compatible with 100 MB and 250 MB disks
|
| |
|
POWER REQUIREMENTS:
|
115/230V AC +/- 10%, 2.5/1.25 A, 50/60 Hz nominal 300 VA max.
|
| |
|
Paper type:
|
Pre-grided thermal paper, Z-fold, 8.5" x 11" (US letter) or 210 mm x 300 mm (A4);
|
|
Printout device:
|
216 mm thermal dot array;
|
|
Paper speed:
|
10, 25, 50 mm/sec. +/- 2%;
|
|
Printing options:
|
Multiple format ECG waveform and alphanumeric printing, diagnostic quality;
|
|
Gain:
|
5, 10, 20 mm/mV +/- 5%;
|
|
Frequency response:
|
0.05 Hz to 150 Hz +/- 10%;
|
|
Muscle artifact filter:
|
User-selectable: Off, 20 Hz or 40 Hz; Baseline filter User-selectable: Off, 0.12 Hz or STABLE (meets or exceeds all 1990 AHA recommendations);
|
|
Calibration signal:
|
1.0 mV for 200 msec. +/- 5%
|
| |
|
Heart rate computation:
|
8-beat average, 30 bpm to 250 bpm, updated every second
|
|
ST measurements:
|
20 sec. moving window with 10 sec. updates to the display and to reports;
|
|
ST parameters:
|
Level, slope for 12 leads simultaneously;
|
|
QRS detection channels:
|
II, V2 and V5 with switching;
|
|
Ectopic detection:
|
Ventricular arrythmias, unclassified;
|
|
ST reference:
|
User-selectable: E-point (QRS onset), J-Junction (QRS offset), ST point (0 msec. to 100 msec., 5 msec. interval) with display/printout annotation
|
| |
|
Preprogrammed:
|
Manual, Bruce, Modified Bruce, Balke, Naughton, Ellestad, Kattus, Ramp, Astrand, Modified Astrand, Pharmacological;
|
|
User-defined:
|
20 protocols with up to 30 stages each plus; Warmup and Recovery phases
|
| |
|
Dimensions:
|
1.3" x 4.3" x 3.8" (3.3 cm x 10.92 cm x 9.65 cm);
|
|
Weight:
|
<1 lb. (0.5 kg);
|
|
Frequency response:
|
0.05 Hz to 150 Hz, +10% to -30% (maximum);
|
|
Defibrillation protection:
|
To 360 J, 5 kV;
|
|
Common mode rejection:
|
>120 dB at 60 Hz;
|
|
Electrode offset tolerance:
|
+/- 300 mV;
|
|
Line frequency filters:
|
User-selectable 50 Hz, 60 Hz or off;
|
|
Noise:
|
<30 μVpp;
|
|
Pacemaker pulse indication:
|
Per EC13;
|
|
Input channels:
|
Simultaneous acquisition of 8 leads plus 4 derived leads;
|
|
Defibrillation recovery:
|
Maximum recovery time is 8 seconds;
|
|
Patient connection:
|
10 replaceable leads
|
| |
|
During exercise:
|
12-lead or Average Complex with Trend Graphs (HR/BP, Single-lead ST, 12-lead ST, Side by side averages, Overlaid averages, 12-lead averages), Rhythm, Ectopic Beat, Event Identification;
|
|
Final report:
|
User-selectable: Exercise Test Summary in tabular or narrative format; Average Complex Summary (Baseline, Maximum ST, Maximum Exercise); Trended Average Complexes (per minute or per stage); Trend Graphs (HR/BP, RPP, METS, Ectopics, 12 single-lead ST); Stress Report Form;
|
|
Comments:
|
Selectable, user-defined;
|
|
Configurations:
|
Up to 5 user-defined configurations with final reports and protocol selections
|
| |
|
Treadmill:
|
Dedicate a 110V 20 amp plug in. Dedicated serial RS-232C for TM55 or TM65; Bike Dedicated serial RS-232C for Treadmill Burdick supported bicycle ergometer
|
| |
|
Control modes:
|
Automatic or manual; Displays Workload in METs; Phase, Stage and Exercise; Test Time (ETT); Ergometer values for treadmill-grade in percent, speed in mph or kmh; for bike -workload in watts, speed in rpm
|
|
Displays:
|
Workload in METs; Phase, Stage and Exercise; Test Time (ETT); Ergometer values for treadmill-grade in percent, speed in mph or kmh; for bike -workload in watts, speed in rpm
|
| |
|
Operating:
|
Temperature 50°F to 104°F (10°C to 40°C);
|
|
Relative humidity:
|
15% to 95% non-condensing;
|
|
Atmospheric pressure:
|
525 mmHg to 795 mmHg (700 hPa to 1060 hPa)
|
| |
|
CONFORMS TO STANDARDS:
|
EN60601-1-1; IEC 601-2-25; EN60601-1-2; ANSI/AAMI EC11, 1991
|
| |
|
SAFETY:
|
IEC 601-1 Type CF Defibrillator Proof
|
| |
|
WARRANTY:
|
Quest system, 3 years; treadmill, 13 months
|